Products

HM-200

HM200 can display N/S, with the high resolution of 0.1G, it can be used for detecting the residual magnetic of machining, the remanence of workpiece and controlling the magnetic situation of plating materials, also detecting the magnetism of motors. With AC measurement function, it can be applied for detecting AC space magnetic field distribution.

SKU: HM200 Category:

Description

Introduction

HM-200 is a portable AC and DC gaussmeter among the LZ series gaussmeters. It is used for measuring the surface magnetic field of permanent magnets and the working magnetic field of DC motor and magnetic separators,HM200 can display N/S, with the high resolution of 0.1G, it can be used for detecting the residual magnetic of machining, the remanence of workpiece and controlling the magnetic situation of plating materials, also detecting the magnetism of motors. With AC measurement function, it can be applied for detecting AC space magnetic field distribution.

Product Overviews

General Features

Portable, convenient to handle and store, for on site measurement

Automatically shut down time can be set, continuously use for more than 30 hours

DC/AC magnetic field

DC/AC maximum value hold

DC field polarity judge function (N/S)

Probe automatic zero, relative measurement, upper and lower limit alarm sorting

Units: G, T, Oe, A/m

Transverse probe with independent calibration value

 

Specifications

Power four AA batteries (alkaline battery is recommended
Display 3digital LCD display
DC Measurement Range 30kG, 3kG, 300G
Resolution 10G, 1G, 0.1G
Accuracy (2% of reading +0.1% of range) (B<20kG, probe temperature15-35C)
Temperature coefficient Maximum (0.06% of reading+0.01G)/C,( probe temperature 15-35C)
Display update rate 2.5 rgd/s
AC Measurement Range 3kG, 300G
Resolution 1G, 0.1G
Bandwidth 10Hz-10kHz (sine magnetic field)
Accuracy (5% of reading +0.2% of range) (effective measurement range, 1kHz sine magnetic field)
Effective measurement range >0.33% range
Rated measurement range >1% range
zero deviation <0.2% range
Temperature coefficient Maximum (0.06% of reading)/C(probe temperature 15-35C)
Display update rate 2.5 rgd/s
Application

Environment

Instrument 15-35C (rated accuracy)
35-45C (accuracy reduced by 0.2%)
Probe 15-35C (rated accuracy)
35-45C(accuracy reduced by 1%)
Humidity 35~75%
Interface 5-core probe socket
Dimension 17*39*142mm (LWT)

 

 Probes: transverse probe and axial probe

Additional information

Weight 1 kg
Dimensions 30 × 25 × 10 cm