Products
HM-6200
It can measure the total magnetic flux,So as to get the values of flux density and magnetic field strength,The HM6200 fluxmeter is very valuable in magnetizing machine, manual and automatic magnet detection and classification, hysteresis loop measurement system,Hall probe can be connected to test the magnetic field strength more accurately
Description
The third generation HM6200 intelligent high precision fluxmeter/gauss meter is controlled by microprocessor,Using the principle of electronic integration, automatic digital suppression adjustment drift system,An instrument that displays the magnitude of magnetic flux and the intensity of magnetic field measured with a color display screen,
It can measure the total magnetic flux,So as to get the values of flux density and magnetic field strength,TheHM6200 fluxmeter is very valuable in magnetizing machine, manual and automatic magnet detection and classification, hysteresis loop measurement system,Hall probe can be connected to test the magnetic field strength more accurately,
It’s all digital multiplex.Provides digital rs-232, USB and relay signal output,Equipped with various helmholtz coil and single coil.External control switch can be connected to reset..External control switch can be connected to reset.
* reset time can be set automatically reset * digital automatic adjustment drift
Measurement specification |
measured parameters
|
||
Coil input | 1 | Magnetic flux measurement accuracy: | readable±0.50%±0.05% range |
Probe input | 1 | Magnetic intensity measurement accuracy: | readable±0.50%±0.05% range |
Displays refresh rate | 3Times/SEC. | Magnetic flux (F) measurement range | 20mWb;200mWb;2000mWb |
Coil compatibility | HM Coil or homemade coil | Magnetic intensity (B/M) measurement range: | 0-3.0T(30kG) |
Probe compatibility | HM Series probe | Magnetic moment (M) : | mWb.cm |
The probe function | Automatic identification and zero correction | Flux measurement resolution | 1µWb,10µWb,100µWb |
Coil connector | diameters 3.8mmplug-in | Resolution of magnetic intensity measurement: | 0.1GS (0.01mT) |
Head connector | 15D (DB15) | temperature coefficient
|
readable±0.03%/oC |
Communication interface 232 function baud rate:19200,57600,115200 | input impedance: | 1KΩ,10KΩ,100KΩ,1000KΩ | |
Connector: 9-pin type D (DB9), DCE setting (connect to PC directly by cable) optional: remote control system or magnetic field control system (aviation or column connector) |
1:Power interface and switch
2:Industrial control interface
3:RS-232
4:Offset adjustment knob
5:Hall probe interface
6: Industrial control terminal
7: Magnetic flux line through
The HM6200 fluxmeter USES 320*240 color LCD screen.In normal operation, the first line of the monitor displays the currently selected range (2mWb;20 MWB;200 MWB;2000mWb) and measurement mode (F: magnetic flux measurement;B: magnetic field strength;M: magnetization);The second line shows the status prompt for holding and clearing zeros;The third line shows the size of the measured value. The measured value has different meanings and units according to the selected measurement mode, as shown in the following table:
Measurement model |
Meaning of measured value |
Optional unit |
F |
Magnitude of magnetic flux |
mWb,KMT |
B |
Magnetic field intensity |
mT,Gs,A/m |
M |
Magnetic moment |
mWb.cm |
Specification and model of magnetic flux meter Helmholtz coil | specifications | Homogeneous area | uniformity | constant | |
Outer diameter:100mm
highly:80mm resistance:327Ω |
Φ25mm | 1% | 0.004654cm | ||
Outer diameter:200mm
highly:133mm resistance:654Ω |
Φ50mm | 1% | 0.009308cm | ||
Outer diameter:300mm
highly:189mm resistance:981Ω |
Φ75mm | 1% | 0.013962cm | ||
Outer diameter:400mm
highly:246mm resistance:1308Ω |
Φ100mm | 1% | 0.018256cm |
Adapter coil
The lamp (additional purchase) Used to remind the magnetic field of the upper limit, the lower limit and the magnetic field under normal state
|
||||
|
Data reading software
The standard:Software installed in the computer to communicate with gauss meter, which can be used to save, export and draw test data The default functions of the two switches in the external controller are data uploading and zero clearing. If there are special requirements, special instructions can be given before the equipment is shipped.When reading, select “HM6200” magnetic flux manual or “HM6200” magnetic HFT in “instrument model” of the reading software, and the software will switch to manual control mode. |
|||
|
External controller (additional purchase)
The fluxmeter can be used to upload and reset data by connecting external switches.Simplified equipment operation during work piece inspection, external controller can be selected in the attachment foot (hand) controller.The external controller is connected to the fluxmeter through the rs-232 interface on the left side of the back panel. The cable length can be customized according to the actual requirements.The external controller is a plug and play device and does not need to be configured on the fluxmeter. |
|||
|
Additional information
Weight | 5.2 kg |
---|---|
Dimensions | 44 × 40 × 25 cm |