Products
FE-30SST
The design of hardware circuit is according to Chinese industry standard(GB / 3655-2000), through real-time sampling for the secondary induction voltage V and primary exciting current I to compensate air to below 1.0mV. It adopts the wave feedback technology, to ensure that the sinusoidal waveform of magnetic flux, compared with similar products, the measuring data will be more accurate.
Description
FE-30SST Silicon Steel Sheet Iron Loss Tester Magnetic Core Lossmeter
Product Description
FE-30SST Silicon Steel Sheet Iron Loss Meter adopts SCM technique and analog electronics technique, the magnetic circuit of permeameter adopts low loss silicon steel iron core, it is a small tester for measuring the characteristics of silicon steel sheet, met the requirements of tracing the brand of silicon steel sheet.
The design of hardware circuit is according to Chinese industry standard(GB / 3655-2000), through real-time sampling for the secondary induction voltage V and primary exciting current I to compensate air to below 1.0mV. It adopts the wave feedback technology, to ensure that the sinusoidal waveform of magnetic flux, compared with similar products, the measuring data will be more accurate.
Instrument features
Panel direct operation, stand-alone can complete the set point of the test
Measurement of loss ratio of single – chip stator chip P1.5 / 50 by open – circuit magnetometer
The closed-loop measurement can lock the magnetic flux Bm test over the total loss
The closed-circuit measurement can be locked by the magnetic flux Hm to test the magnetic induction
Closed magnetic flux meter really made a zero flux compensation
Closed – circuit magnetometer for accurate measurement of oriented, non – oriented silicon steel
The connection computer can perform a variety of curve tests
Software features
The software system supports WinXP / NT / 7 and other mainstream operating systemsThe test process is automatic, no manual operation (only need to put the sample)(B) to determine the magnetic field (H) value, you can also lock the magnetic field (H) to measure the magnetic induction (B) valueReal-time acquisition of waveforms during testing and discontinuation of operation at any timeThe software uses a database file format that supports direct printing and exporting the results to an Excel spreadsheetCan be directly on the data to save, delete, clear and other operationsCan be multi-point measurement, and the data can be set up clusterThrough the multi-point cluster, can display hysteresis loop, magnetization curve, permeability curve and loss curveSupport a variety of models of the printer, the test report can be directly hit The test report can generate JPG, Bmp file format, and can send a report file via email.
Instrument Parameter
Name | Parameter |
Power Supply | 220V ±20V/0.5A,50HZ |
Output Power | 15V/1.5A |
Test Frequency | 50Hz, 60Hz free choice |
Test model | Lock B Measure H and Lock H Measure B values for both modes |
Bm setting range | 500mT~1900mT |
Hm setting range | 100A/m~10000A/m |
Accuracy rating | PS Accuracy 2.0%, Bm Accuracy 1.0% |
Dimension | 280mm*90mm*300mm |
Demo requirements | Length: more than 100mm, width: 20mm ~ 30mm
Thickness: 0.1mm to 0.65mm Requires sample formation, strictly speaking, need to sample the annealing process to trace the national standard. |
Basic configuration
No. | Name of part | Model | Quantity |
1 | Test Host | FE-30SST | 1 |
2 | Monolithic silicon steel magnetic conductivity meter | FE-30SST | 1 |
3 | Cable | Five-core air connector | 1 |
4 | Manual CD | FE-30SST | 1 |
5 | Two pieces of silicon steel samples (manufacturers testing) | Grade: 470 | 1 |
Optional
No. | Name of part | Model |
1 | Measuring software | FE-2120 |
2 | Test computer | Lenovo brand computer B4500 (subject to contract) |
3 | Printers | HP laser black and white printer (subject to contract) |
4 | Open probe | Sample width and length have a certain requirement (20mmX40mm) |
5 | Third-party approved silicon steel samples (16) | Metrology institutes in Zhejiang Province issued |
Test Graph: